Publications
Filter
-
Technical Document 35 Discharges of I-131 and Lu-177 from Oslo hospitals before and after installing a decay tank, and levels in seaweed in the Oslofjord
Levels in seaweed at the discharge point in inner Oslofjord were measurable, even though they were low, after discharges of I-131 and Lu-177 from two Oslo hospitals.
-
StrålevernRapport 2:2017 Ionising radiation metrology infrastructure in Europe
A survey in European counties’ regulations about IR metrology regulations and calibration requirements in radiation protection legislation.
-
Teknisk dokument 8 Air kerma measurements with Landauer nanoDots in Cs-137 and Co-60 beams
Rapporten omhandler målinger av luftkerma i luft med Landauer nanoDots (Landauer, Inc., Greendwood, IL) i Cs-137 og Co-60 felt. Den etablerer gjennom målinger og simuleringer at nanoDots trenger ekstra, energi-tilpasset buildup for målinger av luftkerma i luft ved disse strålingsvalitetene.
-
Teknisk dokument 7 Absorbed doses to water for x-ray dosimetry on a PXI X-RAD 225
For a tube potential of 225 kV, current of 10 mA and source-to-surface distance (SSD) of 50 cm, the absorbed dose rate to water on the central field axis at a depth of 2 g/cm2 of water in the 21.0 cm×21.0 cm×11.5 cm water phantom was (1.16±0.02) Gy/min.
-
Nordic dosimetric capabilities
En oversikt over dosimetrikompetanse, behov og planer i de nordiske land.